Product Description

Sizes of Temperature Humidity Chambers

Model

Volume(L)

Temperature Range
(oC)

Rated Power

Internal Dimension(mm) D x W x H

External Dimension(mm) D x W x H

LRHS-101B-LW

101

A:-40oC∼150oC
B:-55oC∼150oC
C:-65oC∼150oC

17.0kw

450 x 500 x 450

1600 x 1480 x 2000

LRHS-150B-LW

150

22.0kw

500 x 600 x 500

1650 x 1580 x 2100

LRHS-234B-LW

234

30.0kw

750 x 600 x 500 

2200 x 1580 x 2200

LRHS-512B-LW

512

50.0kw

1000 x 900 x 550

2650 x 2050 x 2250

LRHS-753B-LW

753

75.0kw

1300 x 1050 x 550

3100 x 2200 x 2250

LRHS-1000B-LW 

1000

100.0kw

1390 x 1200 x 600

3400 x 2400 x 2300

Basic Specification

Performance

Temperature Shock Range

A:-40oC∼150oC  B:-55oC∼150oC    C:-65oC∼150oC

Temperature Fluctuation

±1oC  

Ramp Rate

Heating: RT∼150oC  <=25min
Cooling: RT∼ -55oC  <=45min
Note: can be customized as per requirement

Temperature Recovery Time

<=5min

Temperature Change Time

<=15s

Material

External Casing

Cold-rolled A3 (Q235) steel plates with plastics painted. 

Interior Working Room

Stainless Steel Plates SUS304 with miror surface

Heat Isolation

Polyurethane foams and fine glass fibres

Illumination Lamp

Philips® one in per standard test chamber.

Casters

4pcs, with PU covered for protection. Height adjustable and direction universal. 

Door

Manually operated sliding door with unloCk button

Heating & Cooling System

Heater

Steipped Nickel-chromium Wire Heater

Air Blower

Anti-temperature stainless fan blades with low noise

Compressor

Hermetically-sealed Techmesh® or semi-sealed Bitzer® or GEA® or equivalent compressors

Cooling Method

Single or dual cooling (Air-cooled or Water-cooled)

Refrigerant

R404A or/and R23 (eco-friendly)

Condenser

Plate-type heat exchanger

Control System

Controller

7" Programmable colorful touch-screen display
LCD English Display
MicroComputer integrated controller

Functions

Data Record (Max 600 days x 24hrs), History Curve, USB data download & upload, Remote controlling through PC, Automatic protection alarms, etc.

Program

1∼1000 programs, 1∼999 steps for per program, Max 99hrs 59mins for per program. 10 programs can be interlinked.

Temperature Sensor

Platinum Resistance.PT100Ω

Running Model

Constant/Fix-value Model and Programmable Model

I/O Interface

Communicate port (RS485) to connect test chamber to a PC

Standard Configuration

Test Chamber x 1, Specimen Holder Shelf x 2, φ100mm Cable Access Port x 1 with 1 soft plug, EMO x 1, Tri-color Indicator x 1, etc.

Safety Protection

Power leakage protection/Compressor over-heat, over-load or over-current protection/Dry burning protection.

Voltage

380V/440V±10%, 50Hz/60Hz or as per requirement

Ambient Enviromental for application

5oC∼+30oC <= 85%R.H

Test Standards (not limited to)

IEC60068-2-1

IEC60068-2-2

IEC60068-2-78

IEC60068-2-30